IRMS data
Laboratory Performance RSD
(Based on 6 13 C values of n-Alkanes)
C22
In compound specific 8 13 C- isotope ratio mass spectrometry, one criterion
commonly used is 1 %o (per mil PDB). This is partly given by the instrumental
measuring uncertainty of around 0.5%o. This means, if one sample’s measuring
is within the range of 1 per mil of another sample, the 8 13 C- content of the mate
rial can be considered the same, or cannot be further distinguished.
Apart from the laboratory performance where the RSD is used to describe the
repeatability (Fig. 13), the defining feature here is the absolute difference of
S 13 C with a threshold of 1 %o.
The NFI followed a straightforward approach and performed analysis, in dupli
cates, on the solutions as received without fractionation or similar preparative
techniques.
[613C] Means of sample #1 {n=1), #10 {n=2)