Table 16: Comparison of quantification of chlordanes [ng/g dw] in sediments by ECNI (two instruments) 
and EI-MS/MS. The fraction of the Florisil clean-up is assigned, where the compound is present. 
Sample 
Method 
Hepta- 
tram- 
cis- 
tram- 
cis- 
chlor 
Chlordane 
Chlordane 
Nonachlor 
Nonachlor 
Florisil fraction 
FI 
F2 
F2 
F1+F2 
F2 
721 (2004) 
EI-MS/MS 
0.054 
0.153 
0.116 
0.13 
0.039 
ECNI, 1200L 
0.061 
0.142 
0.141 
0.10 
0.034 
ECNI, HP 5989B 
0.064 
0.135 
0.126 
0.10 
0.029 
Average ± STD 
0.056± 
0.143± 
0.128± 
0.110± 
0.034± 
0.005 
0.007 
0.010 
0.014 
0.004 
ODER (2004) 
EI-MS/MS 
0.056 
0.066 
0.083 
0.12 
0.021 
ECNI, 1200L 
0.036 
0.078 
0.044 
0.10 
0.026 
ECNI, HP 5989B 
0.042 
0.059 
0.060 
0.09 
0.017 
Average ± STD 
0.045± 
0.068± 
0.062± 
0.103± 
0.021± 
0.008 
0.008 
0.016 
0.012 
0.004 
715 (2004) 
EI-MS/MS 
<0.4 
0.047 
0.027 
0.60 
0.023 
ECNI, 1200L 
<1.3 
0.049 
0.038 
0.50 
0.024 
ECNI, HP 5989B 
<1.6 
0.042 
0.031 
0.50 
0.021 
Average ± STD 
0.046± 
0.032± 
0.53± 
0.023± 
0.003 
0.005 
0.05 
0.001 
RÜDEN 
EI-MS/MS 
<0.4 
0.018 
0.023 
0.20 
0.011 
(2004) 
ECNI, 1200L 
<1.3 
0.014 
0.025 
0.16 
0.013 
ECNI, HP 5989B 
<1.6 
0.006 
0.016 
0.15 
0.006 
Average ± STD 
0.013± 
0.021± 
0.17± 
0.010± 
0.005 
0.004 
0.02 
0.003 
ECKFBU 
EI-MS/MS 
<0.4 
0.047 
0.033 
0.17 
0.032 
(2004) 
ECNI, 1200L 
<1.3 
0.044 
0.054 
0.15 
0.027 
ECNI, HP 5989B 
<1.6 
0.036 
0.035 
0.14 
0.019 
Average ± STD 
0.042± 
0.041± 
0.15± 
0.026± 
0.005 
0.009 
0.01 
0.005 
STD: Standard deviation 
5.3.7 Method validation and quality control 
5.3.7.1 Limits of detection and quantification 
Typically, limits of detection (LOD, signal-to-noise ratio 3:1) for s- or mPCAs in fish 
liver were about 1 ng/g and limits of quantification (LOQ, signal-to-noise ratio 10:1) 
were ca. 3 ng/g. Chlordanes could be identified down to 1-3 pg/g and quantified from 
ca. 2-10 pg/g and onwards. 
Single PCA congener groups in sediments were detectable down to ca. 5-50 ng/g 
depending on their structure. Limits of quantification were about 3-5 times higher. 
Total PCAs in sediments could be quantified with EI-MS/MS down to 2-8 ng/g.
	        
Waiting...

Note to user

Dear user,

In response to current developments in the web technology used by the Goobi viewer, the software no longer supports your browser.

Please use one of the following browsers to display this page correctly.

Thank you.