Environ Sci Pollui Res (2015) 22:19887-19895
19893
Springer
Internal Standards
Fig. 5 Recovery rates of internal laboratory standards on deployed
samplers from Fehmarn and Heligoland (n = 2 each) and their
corresponding lab (n=3) and transport blanks (h=2)
Removing silicone oligomers from sample extracts
by HPLC-SEC
Measurements of PDMS extracts with TXRF showed co
extraction of silicone oligomers (see “Optimization of the sili
con rubber pre-cleaning”). Hence, extracts had to be further
purified to avoid coating of instrumental parts. Smedes and
Booij (2012) recommend extract purification with C-l 8 bonded
silica cartridges, which is suitable for methanol or acetonitrile/
methanol as solvent. Due to the fact that the ideal solvent for
PDMS sampler extraction with ASE was found to be non-polar
(“Optimization of the organic solvent for ASE extraction”),
alternative purification steps needed to be performed.
Shahpoury and Hageman (2013) used SEC to remove silicone
oligomers from sample extracts only for PAH analysis. HPLC-
SEC can be set up with non-polar solvents and hence no solvent
transfer from non-polar to polar and back is necessary, which
can potentially result in lower analyte recoveries by using C-l8
silica cartridges. In this study, HPLC-SEC was tested for the
separation of PAH and CHCs from silicone oligomers using a
mixture of hexane-dichloromethane as eluent in order to keep
the same extraction solvents during the entire extraction proce
dure. Monitoring the resulting SEC fractions with TXRF
showed that the majority of silicone oligomers (mean 98.9 %)
were present in the first fraction (0-10 min). Less than 1 %
(average 0.3 %) of the original silicone oligomer content was
left in the second fraction (10-28 min), in which the target
CHCs and PAHs were eluted. HPLC-SEC is restricted to a
certain amount of residual oligomers, e.g. if the pre-cleaning
step to remove residual oligomers is skipped, the burden for a
semi-preparative column to separate target compounds from
oligomers is too high. HPLC-SEC thus proved to be a very
efficient purification method to remove co-extracted silicone
oligomers showing the additional advantage that no solvent
exchange is necessary.
Additionally to HPLC-SEC, the fast monitoring of the sil
icone content in each (purified) extract by TXRF avoids ana
lytical interferences and coating of instrument parts (e.g. GC
liner and column).
Field samples
Recovery rates
The whole set of deployed sampler, transport and laboratory
blanks was extracted and purified as described in “Sample
extraction”. Internal laboratory standard recoveries represent
the quality of the extraction and purification procedure and
ranged from 83-114 % for CHCs and PAHs (Fig. 5). De
ployed PDMS samplers from Heligoland and Fehmam also
showed good IS recovery rates for CHCs in the range from 85
to 126 % (Fig. 5), which is within the range of the
Analyt
OOU-WZ-J>-<|-£9<COI =
<< i<u-CLCOQ;mmc\j<:fc
CL I — CO ^
. o o m
Analyt
Fig. 6 a Fits of the proportionality constant B by nonlinear least square regression, b, c Calculated concentrations of CHCs and PAHs in water from
deployed sampler (n=2) of Fehmarn (Baltic Sea) and Heligoland (North Sea)