20
Oil Spill identification - Round Robin 20045
LASEM has diluted the extracts a factor 2 and diluted 100pJ of the
source samples in 1 ml DCM. The injection volume is 1 pi. As a result
the source samples should give about a 10 times higher signal than the
spill samples. In annex IV the areas of m/z 216 are shown. From the
table the mean and RSD has been calculated:
extract I
extract II
Source I
Source II
m/z=216
mean
RSD in %
mean
RSD in %
mean
RSD in %
mean
RSD in %
2M-fluoranthene
23055
8.4
23079
9.6
67871
84.8
83779
23.8
benzo(a)fluore
ne
60827
7.4
58517
2.9
233463
64.0
277480
21.1
benzo(b)fluoren
e
24226
11.4
23276
4.5
71067
53.6
69110
42.8
2M-pyrene
14071
21.5
14846
4.7
135941
77.2
177571
19.8
4M-pyrene
29785
11.6
30783
3.6
216228
52.9
306144
9.5
1 M-pyrene
18138
8.2
17829
1.6
106854
49.6
150598
8.4
Table 1. The source results show much higher RSD's than the extract results.
Mean and RDS calculated from annex
IV of the LASEM report
From the table with the analytical methods (see 7.2 of this report) it
can be seen that LASEM uses the lowest injection temperature of all
participants. This subject is discussed in 2.3 Cedre.
It can also be see that only LASEM uses an ion trap for the biomarker
analysis. Petrobas uses the same instrument for the PAH analysis, but
uses a quadrupole instrument for the biomarkers.
An ion trap has to reduce the total number of ions in the source to
prevent chemical ionization reactions and to be able to filter the ions
properly. Therefore the number of ions is measured before each scan
and only a limited number is entering the source. It is therefore possible
that for the samples of RR2005 the number of biomarker ions
compared to the total number of ions is so small that it is difficult to
analyze them correctly with an ion trap.
Note: An email to Petrobas with a request for the reason why the ion
trap is not used for biomarkers has not been answered before finishing
the report.
Advise:
Optimize the injection concentration for bilge samples by
means of a calibration line
Validate the instrument by means of 7 to 8 injections and
calculate the relative st dev of (some of the) the ratios. They
should not be much higher that 5%.
Analyze the same samples on a GC-MS with a quadrupole
detector when possible and compare the results.
2.4 LVA
Contact: Rita Skolmeistere